Comparison of Stresses in Four Modular Total Knee Arthroplasty Prosthesis Designs

Comparison of Stresses in Four Modular Total Knee Arthroplasty Prosthesis Designs

Ahilan Anantha Krishnan, Rupesh Ghyar, Bhallamudi Ravi
Copyright: © 2016 |Pages: 16
DOI: 10.4018/IJBCE.2016070101
OnDemand:
(Individual Articles)
Available
$37.50
No Current Special Offers
TOTAL SAVINGS: $37.50

Abstract

The current study, compares the mechanical performance of four modular TKA prostheses based on von Mises stress distribution in the tibial insert. Three-dimensional finite element (FE) models of a cruciate retaining type modular prosthesis and three posterior stabilized (PS) type modular prostheses namely: anterior slide, modular post and double cam, were developed. A compressive load of 2600 N was applied to the FE models at flexion angles of 00, 150, 300, 600 and 900. Von Mises stress was evaluated on all the modular parts of the prostheses and compared with the yield strength of the corresponding material. Von Mises stress in all the parts were below the yield strength of their corresponding material except for tibial insert of anterior slide design at high flexion angle. Von Mises stress above the yield strength in the tibial insert of anterior slide design, was due to edge loading in the post and it demonstrates the likelihood of mechanical failure by delamination type of wear.
Article Preview
Top

Introduction

The major reasons for the failure and revision of Total knee arthoplasty (TKA) are: polyethylene wear, aseptic loosening, instability and infection. Labek et al. (2011) did a systematic review of national registers and showed that, 1.26 total knee revisions were done per 100 observed component years. Sadoghi et al. (2013) reviewed the worldwide Arthroplasty registers and reported that reasons for TKA revision as: aseptic loosening (29.8%), septic loosening (14.8%), pain (9.5%), wear (8.2%) and others. Fehring et al. (2001) on a study about reasons for early revision (<5 years) on 279 patients found the reasons were: infection (38%), instability (27%), patellofemoral problems (8%) and wear (7%). Narkbunnam and Chareancholvanich (2012) studied the causes for revision in 189 patients and found that the reasons to be aseptic loosening (52%), polyethylene wear (43%) and infection (31%).

Aseptic loosening of the prosthesis is said to be caused by both biological and mechanical responses. Aseptic loosening occurs due to osteolysis, micro-motion of: bone – cement – prosthesis interface and stress shielding (Grewal et al., 1992; Ingham & Fisher, 2000; Lonner et al., 2001). Of which, osteolysis is a direct biological consequence of wear debris from UHMWPE, causing adverse cellular reactions and inflammation (Purdue et al., 2006; Burton et al., 2012). Stress shielding occurs after TKA process, because of high relative stiffness between prosthesis and bone thereby causing major load transfer through the prosthesis. Thus, this leads to weakening and resorption of bone, resulting in loss of bone – prosthesis support, further leading to component loosening (Lonner et al., 2001). Wear in the tibial insert made of UHMWPE is another major reason for revision surgery and it is a multifactorial issue comprising of the patient, prosthesis and surgical factors. Wear in the tibial insert is influenced by: activity level of the patient, manufacturing of tibial insert, design, alignment and ligament balance during surgery (Mcgloughlin & Kavanagh, 2000; McEwen et al., 2005; Naudie et al., 2007).

Under real-time loading, cyclic stresses above the yield strength in the tibiofemoral interface causes: pitting, delamination, change in the microstructure and degradation of UHMWPE (Landy & Walker, 1988; Edidin et al., 1999). When the von Mises stress exceeds the yield strength of the material, plastic deformation occurs. This plastic strain gets accumulated causing a fatigue crack formation and an eventual failure of polyethylene surface layer by delamination (fatigue wear) (Cooper et al., 1993; Kennedy et al., 2000). Von Mises stress which is a measure of delamination type of wear (Kennedy et al., 2000; Cho et al., 2004) can be easily found by FE analysis, unlike experimental testing. Failure can occur in components other than tibial insert due to lack of bone support, thus causing high stresses above the yield strength. For instance, Huang et al., (1999) reported fracture of the femoral component in a patient due to loss of osseous support. Tibial tray failure was also reported in the literature and they were mainly related to design, mal-alignment and inadequate bone support (Boran et al., 2005; Chatterji et al., 2005).

Complete Article List

Search this Journal:
Reset
Volume 12: 1 Issue (2024): Forthcoming, Available for Pre-Order
Volume 11: 2 Issues (2022): 1 Released, 1 Forthcoming
Volume 10: 2 Issues (2021)
Volume 9: 2 Issues (2020)
Volume 8: 2 Issues (2019)
Volume 7: 2 Issues (2018)
Volume 6: 2 Issues (2017)
Volume 5: 2 Issues (2016)
Volume 4: 2 Issues (2015)
Volume 3: 2 Issues (2014)
Volume 2: 2 Issues (2013)
Volume 1: 2 Issues (2012)
View Complete Journal Contents Listing